18.09.2012 Views

Stratos®Pro A4... PH User Manual

Stratos®Pro A4... PH User Manual

Stratos®Pro A4... PH User Manual

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

5<br />

Menu item Action Choices<br />

Time averaging filter Enter value using ��<br />

�� keys.<br />

Press enter to confirm.<br />

Configuration<br />

Time averaging filter<br />

To smoothen the current output, a low-pass filter with adjustable filter<br />

time constant can be switched on. When there is a jump at the input<br />

(100 %), the output level is at 63 % after the time interval has been<br />

reached. The time interval can be set from 0 to 120 sec. If the time<br />

interval is set to 0 sec, the current output directly follows the input.<br />

Please note:<br />

The filter only acts on the current output, not on the display, the limit<br />

values, or the controller!<br />

During HOLD the filter is not applied. This prevents a jump at the<br />

output.<br />

pH<br />

Temp<br />

Display / Controller<br />

Time interval 0...120 s<br />

Time interval 0...120 s<br />

0/4-20 mA<br />

81