ORP Calibration<br />

Display Action Remark<br />

Select ORP calibration,<br />

proceed with enter<br />

Remove the sensor<br />

and temperature<br />

probe, clean them, and<br />

immerse them in the<br />

redox buffer.<br />

Enter setpoint value for<br />

redox buffer.<br />

Press enter to proceed.<br />

The ORP delta value is<br />

displayed (based on<br />

25°C).<br />

Sensoface is active.<br />

Press enter to proceed.<br />

To repeat calibration:<br />

Select REPEAT.<br />

To exit calibration:<br />

Select MEAS, then press<br />

enter<br />

Display (3 sec)<br />

Now the device is in<br />

HOLD mode.<br />

After end of calibration,<br />

the outputs remain<br />

in HOLD mode<br />

for a short time.<br />

135