-U1- Specifiable Buffer Set<br />

You can specify a buffer set with 2 buffer solutions in the temperature<br />

range of 0 ... 95 °C, step width: 5 °C.<br />

To do so, select buffer set -U1- in the configuration menu.<br />

As delivered, the Ingold technical buffer solutions pH 4.01 / 7.00 are<br />

stored as buffer set and can be edited.<br />

Conditions for the specifiable buffer set:<br />

• All values must lie in the range pH 0 ... 14<br />

• Maximum difference between two adjacent pH values<br />

(5 °C step width) of the same buffer solution: pH 0.25<br />

• The values of buffer solution 1 must be lower than those of<br />

buffer solution 2:<br />

The difference between values for identical temperatures must be<br />

greater than 2 pH units.<br />

Faulty entries are indicated in measuring mode by the<br />

“FAIL BUFFERSET -U1-“ message.<br />

The 25 °C value is always used for buffer display during calibration.<br />