Keypad Keys: meas, menu, info, 4 cursor keys, enter<br />

Specifications<br />

HART communication HART version 6<br />

Digital communication by FSK modulation of output current 1<br />

Device identification, measured values, status and messages,<br />

parameter setting, calibration, records<br />

Conditions Output current ≥ 3.8 mA and load resistance ≥ 250 Ω<br />

IrDA interface Infrared interface for firmware update<br />

FDA 21 CFR Part 11 Access control by editable passcodes<br />

Logbook entry and flag via HART in the case of configuration changes<br />

Message and logbook entry when enclosure is opened<br />

Diagnostics functions<br />

Calibration data Calibration date, zero, slope, response time<br />

Device self-test Display test, automatic memory test (RAM, FLASH, EEPROM),<br />

module test<br />

Logbook 100 events with date and time<br />

Extended logbook (TAN) Audit Trail: 200 events with date and time<br />

Service functions<br />

Sensor monitor Display of direct sensor signals<br />

Current source Current specifiable for output 1 and 2 (00.00 ... 22.00 mA)<br />

Relay test <strong>Manual</strong> control of the four switching contacts<br />

<strong>Manual</strong> controller Controller output entered directly (start of control process)<br />

IrDA Activating the IrDA function<br />

Passcodes Assigning passcodes for menu access<br />

Factory setting Resetting all parameters to factory setting<br />

TAN Enabling optionally available additional functions<br />

Data retention Parameters, calibration data, logbook > 10 years (EEPROM)<br />

157