Menu item Remark<br />

Controller<br />

characteristic<br />

Relay 1<br />

Setpoint<br />

+100 %<br />

-100 %<br />

Controller output<br />

Yp [%]<br />

Neutral zone Yp=0<br />

Deviation<br />

Xw<br />

Relay 2<br />

The arrows indicate which<br />

relay (valve) is active:<br />

Relay 2 active<br />

(Meas. value > setpoint)<br />

Relay 2 active<br />

(Meas. value < setpoint)<br />

Service<br />

Controller test (manual specification<br />

of controller output):<br />

This function is used to start up control loops or<br />

check the actuators.<br />

For bumpless changeover to automatic operation<br />

(exiting this function), configure an I-action<br />

component (reset time).<br />

The lower display shows the currently adjusted<br />

controller output Yp.<br />

Specify new value for controller output Yp:<br />

Enter sign and value in the main display using the<br />

����keys. Press enter to confirm.<br />

The new value is taken into the lower display.<br />

Return to Service menu: Press meas.<br />

Return to measurement: Hold meas depressed for<br />

longer than 2 sec.<br />

Controller output –100...0% Relay 2 active<br />

Controller output 0...+100% Relay 1 active<br />

IrDA communication:<br />

Select IRDA using��,<br />

press enter to confirm.<br />

When IrDA communication is active, the device<br />

remains in the HOLD mode for reasons of safety.<br />

Further operation is performed via IrDA.<br />

End communication by pressing meas.<br />

Exception: Firmware update<br />

(must not be interrupted!)<br />