Calibration<br />

Please note:<br />

• All calibration procedures must be performed by trained personnel.<br />

Incorrectly set parameters may go unnoticed, but change the<br />

measuring properties.<br />

• The response time of the sensor and temperature probe is<br />

considerably reduced when the sensor is first moved about<br />

in the buffer solution and then held still.<br />

• The device can only operate properly when the buffer solutions<br />

used correspond to the configured set. Other buffer solutions, even<br />

those with the same nominal values, may demonstrate a different<br />

temperature response. This leads to measurement errors.<br />

When using ISFET sensors or sensors with a zero point other<br />

than pH 7, the nominal zero point must be adjusted each time a new<br />

sensor is connected. This is important if you want to obtain reliable<br />

Sensoface messages. The Sensoface messages issued during all further<br />

calibrations are based on this basic calibration.<br />

122