Stratos®Pro A4... PH User Manual
Stratos®Pro A4... PH User Manual
Stratos®Pro A4... PH User Manual
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
5<br />
Menu item Action Choices<br />
Temp detection<br />
during calibration<br />
Select mode using ��<br />
keys:<br />
AUTO: Measured by<br />
sensor<br />
MAN: Direct input of<br />
temperature, no measurement<br />
(see next step)<br />
EXT: Temperature specified<br />
via current input<br />
(only if TAN E enabled)<br />
Press enter to confirm.<br />
(<strong>Manual</strong> temperature) Modify digit using ��<br />
keys,<br />
select next digit using<br />
�� keys.<br />
Press enter to confirm.<br />
Calibration mode<br />
(AUTO: Buffer set)<br />
Select CALMODE using<br />
�� keys:<br />
AUTO: Calibration with<br />
Calimatic buffer set<br />
recognition<br />
MAN: <strong>Manual</strong> entry of<br />
buffer solutions<br />
DAT: Input of adjustment<br />
data of premeasured<br />
sensors<br />
Press enter to confirm.<br />
Select buffer set using<br />
�� keys (see buffer<br />
tables for nominal<br />
values).<br />
Press enter to confirm.<br />
Configuration<br />
-00-...-10-,<br />
(-U1-, see Appendix)<br />
Pressing the info key displays<br />
the manufacturer<br />
and nominal values in the<br />
lower line.<br />
65