Specifications<br />

I input (TAN) Current input 0/4 ... 20 mA / 50 Ω for external temperature signal<br />

Start/end of scale Configurable –20 ... 200 °C (-4 ... 392 °F<br />

Characteristic Linear<br />

Measurement error 1.3)<br />

154<br />

< 1% current value + 0.1 mA<br />

HOLD input Galvanically separated (OPTO coupler)<br />

Function Switches device to HOLD mode<br />

Switching voltage 0 ... 2 V (AC/DC) HOLD inactive<br />

10 ... 30 V (AC/DC) HOLD active<br />

CONTROL input Galvanically separated (OPTO coupler)<br />

Function Selecting parameter set A/B or flow measurement<br />

Parameter set A/B Control input 0 ... 2 V (AC/DC)<br />

10 ... 30 V (AC/DC)<br />

FLOW Pulse input for flow measurement 0 ... 100 pulses/s<br />

Pulse amplitude 10 ... 30 V DC<br />

Message via 22 mA, alarm contact or limit contacts<br />

Display 00.0 … 99.9 l/h<br />

Parameter set A<br />

Parameter set B<br />

Output 1 0/4 ... 20 mA, max. 10 V, floating (galv. connected to output 2)<br />

Process variable * pH, ORP or temperature<br />

Characteristic Linear<br />

Overrange * 22 mA in the case of error messages<br />

Output filter * PT 1 filter, time constant 0 ... 120 s<br />

Measurement error 1)<br />

Start/end of scale *<br />

< 0.25 % of current value + 0.025 mA<br />

Configurable within selected range<br />

Minimum span pH 2.00 / 200 mV / 20 K / 36 °F