Menu item Action Choices<br />

CIP / SIP<br />

The following adjustments are possible for ISM sensors :<br />

Cleaning cycle<br />

counter<br />

Cleaning cycles<br />

Sterilization cycle<br />

counter<br />

Sterilization cycles<br />

Select ON or OFF using<br />

�� keys.<br />

Press enter to confirm.<br />

Only with CIP COUNT ON:<br />

Enter value using ��<br />

�� keys.<br />

Press enter to confirm.<br />

Select ON or OFF using<br />

�� keys.<br />

Press enter to confirm.<br />

Only with CIP COUNT ON:<br />

Enter value using ��<br />

�� keys.<br />

Press enter to confirm.<br />

Configuration<br />

ON/OFF<br />

0...9999 CYC<br />

(0000 CYC)<br />

ON/OFF<br />

0...9999 CYC<br />

(0000 CYC)<br />

The cleaning and sterilization cycles are counted to measure the load<br />

on the sensor.<br />

Suitable for biochemical applications (process temp approx. 0...50 °C,<br />

CIP temperature > 55 °C, SIP temperature > 115 °C).<br />

75