Configuration<br />

Parameter set A/B: configurable menu groups<br />

The device provides 2 parameter sets “A“ and “B“. By switching between the parameter<br />

sets you can adapt the device to different measurement situations, for example.<br />

Parameter set “B“ only permits setting of process-related parameters.<br />

Menu group Parameter set A Parameter set B<br />

Sensor selection ---<br />

Current output 1 Current output 1<br />

Current output 2 Current output 2<br />

Compensation Compensation<br />

Control input ---<br />

Alarm mode Alarm mode<br />

Relay outputs Relay outputs<br />

Cleaning ---<br />

Parameter set selection ---<br />

Setting the clock ---<br />

Tag number ---<br />

External switchover of parameter sets A/B<br />

You can switch between parameter sets A and B by applying a signal<br />

to the CONTROL input (parameter setting: ).<br />

Max. 30 V AC/DC<br />

e.g. process control system<br />

Parameter set A active 0...2 V AC/DC<br />

Parameter set B active 10...30 V AC/DC<br />

12<br />

13<br />

input<br />

CONTROL<br />