Data Entry of Premeasured Sensors<br />

The DAT calibration mode must have been preset during<br />

configuration.<br />

You can directly enter the values for slope and asymmetry potential<br />

of a sensor. The values must be known, e.g. determined beforehand in<br />

the laboratory.<br />

Display Action Remark<br />

Select Calibration.<br />

Press enter to proceed.<br />

130<br />

“Data Input“<br />

Ready for calibration.<br />

Hourglass blinks.<br />

Enter asymmetry<br />

potential [mV].<br />

Press enter to proceed.<br />

Enter slope [%].<br />

The device displays<br />

the new slope and<br />

asymmetry potential<br />

(at 25 °C).<br />

Sensoface is active.<br />

Use the arrow keys to<br />

select:<br />

• MEAS (exit)<br />

• REPEAT<br />

Press enter to proceed.<br />

Display (3 sec)<br />

Now the device is in<br />

HOLD mode.<br />

Exit:<br />

HOLD is deactivated<br />

with delay.