Configuration<br />
Adaptive Calibration Timer (ACT)<br />
By issuing a Sensoface message, the adaptive calibration timer<br />
reminds you to calibrate the sensor. After expiration of the ACT interval,<br />
Sensoface is getting “sad”. Pressing the info key shows the text<br />
“OUT OF CAL TIME CALIBRATE SENSOR” which reminds you that a<br />
calibration is due. The ACT interval is either read automatically from<br />
the sensor settings or can be specified manually (max. 9999 days).<br />
Stressing influences (temperature, measurement in extreme ranges)<br />
shorten the timer interval.<br />
The adaptive cal timer is reset after each calibration.<br />
You can configure the current outputs so that a Sensoface message<br />
generates a 22-mA error signal, see page 83.<br />
Menu item Action Choices<br />
Adaptive cal timer<br />
(ACT)<br />
Select using �� keys:<br />
AUTO:<br />
The interval stored in<br />
the ISM sensor is used<br />
(default)<br />
MAN:<br />
The interval is specified<br />
manually (0 ... 9999 days)<br />
Press enter to confirm.<br />