Configuration<br />

5<br />

Menu item Action Choices<br />

Select function of Select using �� keys. PARSET<br />

CONTROL input Press enter to confirm. (selecting parameter<br />

set A/B via signal at<br />

CONTROL input)<br />

Adjust to flow meter With "Flow" selected,<br />

you must adjust the<br />

device to the flow meter<br />

used.<br />

Enter value using arrow<br />

keys.<br />

Press enter to confirm.<br />

Flow<br />

(for connecting a pulseoutput<br />

flow meter)<br />

12000 pulses/liter<br />

In the alarm menu you can configure flow monitoring. When you have<br />

set CONTROL to FLOW, you can specify 2 additional limit values for<br />

maximum and minimum flow.<br />

If the measured value lies outside this range, an alarm message and a<br />

22-mA error signal (if configured) will be generated.<br />

Display<br />

Flow measurement in measuring mode<br />

Display<br />

Flow measurement (sensor monitor)<br />

95