Sensoface<br />

Display Problem Status<br />

Asymmetry<br />

potential and<br />

slope<br />

178<br />

Calibration<br />

timer<br />

Asymmetry potential (zero)<br />

and slope of the sensor are<br />

still okay.<br />

The sensor should be replaced<br />

soon.<br />

Asymmetry potential and<br />

slope of the sensor have<br />

reached values which no longer<br />

ensure proper calibration.<br />

Replace sensor.<br />

Over 80% of the calibration<br />

interval has already past.<br />

The calibration interval has<br />

been exceeded.<br />

Sensor defect Check the sensor and its<br />

connections (see also Error<br />

Messages Err 15 and Err 16).<br />

Response<br />

time<br />

Sensor response time has<br />

increased.<br />

The sensor should be replaced<br />

soon.<br />

To achieve an improvement, clean<br />

the sensor and soak it in buffer.<br />

Sensor response time significantly<br />

increased ( > 72 s, calibration<br />

aborted after 120 s)<br />

Replace sensor.