Auxiliary<br />

P810 Units<br />

This parameter determines whether the Measurement units (P104) are<br />

displayed on the auxiliary line of the display in run mode.<br />

Option<br />

Description<br />

0 = No Measurement units will not be displayed<br />

1 = Yes (Default) Measurement units will be displayed<br />

P811 Alarms Messages<br />

This parameter determines whether notification messages are displayed on<br />

the auxiliary line of the display in run mode when an alarm relay is switched<br />

on or off. The message is in the form “Alarm High ON”, where the ‘High’ is<br />

determined by the setting of the relay Alarm ID (P212, 222, 232, 242, 252).<br />

Option<br />

Description<br />

0 = No (Default) Alarm messages will not be displayed<br />

1 = Yes Alarm messages will be displayed<br />

P812 General Control or Pump Messages<br />

This parameter determines whether notification messages are displayed on<br />

the auxiliary line of the display in run mode when a pump or general control<br />

relay is switched on or off. The message is in the form “General 1 ON”,<br />

where the number displayed is the number of the relay.<br />

Option<br />

Description<br />

0 = No (Default) Pump messages will not be displayed<br />

1 = Yes Pump messages will be displayed<br />

P813 Control Messages<br />

This parameter determines whether notification messages are displayed on<br />

the auxiliary line of the display in run mode when a control relay is switched<br />

on or off. The message is in the form “Time ON”.<br />

Option<br />

Description<br />

0 = No (Default) Control messages will not be displayed<br />

1 = Yes Control messages will be displayed<br />

Page 102