FlowCERT (UL)
UL manual - Pulsar Process Measurement
UL manual - Pulsar Process Measurement
- No tags were found...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Special Devices<br />
Point of Measurement<br />
The transducer must be above the maximum head P704 by at least the near<br />
blanking distance P107.<br />
In the case of a Palmer Bowlus flume the point of head measurement<br />
should be half the value of Dim “A” P710 upstream of the device.<br />
For an H-Flume the head measurement is taken at a point downstream<br />
from the flume entrance as detailed in the table below:<br />
Flume size Point of Measurement<br />
Dim. “A” P710<br />
cm Feet cm inches<br />
15.25 0.5 4.7 1.88<br />
23.00 0.75 6.7 2.69<br />
30.05 1.0 9.1 3.63<br />
45.70 1.5 13.5 5.38<br />
61.00 2.0 17.9 7.19<br />
76.20 2.5 22.5 9.00<br />
91.45 3.0 27.2 10.88<br />
137.15 4.5 40.5 16.19<br />
Page 55