Using the RS232 Serial Interface<br />

The RS232 serial interface is used to communicate between the <strong>FlowCERT</strong><br />

LITE and a PC using the optional Ultra PC and other associated Pulsar<br />

software packages, to obtain information such as data logging and view echo<br />

traces upload, download and save parameter files. In addition it can also be<br />

used to control or obtain information using a standard PC or other computer<br />

base equipment. To do so, the settings for control are as follows: baud rate<br />

19,200, 8 data bits, no parity, 1 stop bit.<br />

The device should be connected, via the serial port, as shown in Chapter 2<br />

Installation.<br />

To use the device remotely, you need to log on to start, and log off when<br />

finished. When logged on, <strong>FlowCERT</strong> LITE will show ‘Remote ON’ on<br />

the display, and “Communicator OFF” when logged off.<br />

All commands should be followed by a carriage return.<br />

The unit will respond either OK (or a value) if the command is accepted, or<br />

NO if it is not.<br />

To log on, send the command<br />

/ACCESS:pppp where pppp is the passcode (P922).<br />

To log off, send the command<br />

/ACCESS:OFF<br />

To read a parameter value, send the command<br />

/Pxxx where xxx is the parameter you wish to read, and the <strong>FlowCERT</strong><br />

LITE will respond with the parameter value.<br />

To set a parameter, send the command<br />

/Pxxx:yy where xxx is the parameter number, and yy is the value you wish<br />

to set it to.<br />

Page 26