U-Throated Flume<br />

ABSOLUTE<br />

If the flow calculation is to be absolute P702 = 1 the flow will be calculated<br />

using the formula: q = (2/3) 1.5 g n 0.5 C u C v C d bh 1.5<br />

Where: q = flowrate<br />

g n = gravitational acceleration (nominal value = 980.66 cm/s 2 ) h<br />

= head C u =<br />

shape coefficient calculated by <strong>FlowCERT</strong> LITE P724<br />

C v = velocity coefficient calculated by <strong>FlowCERT</strong> LITE P721<br />

C d = discharge coeffecient calculated by <strong>FlowCERT</strong> LITE P722<br />

b = throat width P711<br />

RATIOMETRIC<br />

U-Throated Flume<br />

If the flow calculation is to be ratiometric P702 = 2 the flow will be<br />

calculated using the formula:<br />

q= q cal (C v /C vcal )(C d /C dcal )(C u /C ucal )(h/h cal ) 1.5<br />

Where: q = flowrate<br />

q cal = flowrate at maximum head P705<br />

Cv = velocity coefficient calculated by <strong>FlowCERT</strong> LITE P721<br />

Cv cal = velocity coefficient at maximum head<br />

Cd = discharge coeffecient calculated by <strong>FlowCERT</strong> LITE P722<br />

Cd cal = discharge coefficient at maximum head<br />

Cu = shape coefficient P724<br />

Cu cal = shape coefficient at maximum head<br />

h = head<br />

= maximum head P704<br />

h cal<br />