Cable Entry<br />
There are 6 cable gland knock-outs on the base of the wall mount<br />
<strong>FlowCERT</strong> Lite (5 x 0.79" (20mm), 1 x 0.63" (16mm)) and 4 on the rear (4<br />
x 0.73" (18mm)). Select which ones you wish to use, and remove them by<br />
using a circular cutter, such as a tank cutter. Take care not to damage the<br />
circuit board inside while undertaking this. Do not use a hammer, as this<br />
may cause damage to the enclosure.<br />
It is recommended that you use suitable cable glands to ensure that the<br />
ingress rating is maintained and that they be tightened to the manufacturers<br />
recommended settings.<br />
Important Information<br />
All cable glands should be tightened to the manufacturer’s specifications.<br />
The terminal compartment cover screws should be tightened to 1.48lb<br />
in.(2Nm)<br />
Care should be taken not to over tighten the screws.<br />
Terminal Connection Details<br />
The terminal strip is as detailed below. There is also a wiring diagram inside<br />
the terminals access cover.<br />
Care should be taken not to over tighten the screws.<br />
Important Information<br />
All terminal connection screws should be tightened to 4.5in.lbs.<br />
(0.5Nm).<br />
Care should be taken not to over tighten the screws.<br />