Point of Measurement<br />

The transducer must be above the maximum head P704 by at least the near<br />

blanking distance P107.<br />

For Suppressed/Contracted Rectangular, Trapezoidal and V-notch,<br />

weirs, the head is measured upstream at a minimum distance of 3 times<br />

maximum head from the weir plate to ensure the surface of the liquid is not<br />

affected by turbulence or drawdown. (See DRWG. 1)<br />

In the case of a Venturi flume the point of measurement should be 150 mm<br />

upstream from the beginning of the converging section and for a Parshall<br />

flume 2/3 the length of the converging section upstream of the throat<br />

section. See DRWG 2 and 3)<br />