To program the <strong>FlowCERT</strong> LITE for Example 1 V-Notch Weir by using<br />
the Quick Setup menu proceed as follows.<br />
If required access the Program Mode<br />
Key in the passcode 1997 and press ENTER<br />
Using the ‘right’ arrow key go to the Quick Setup menu press ENTER and<br />
as prompted, by the questions, select the relevant option and press ENTER.<br />
Question<br />
Option<br />
PMD Type 1 = Exponent<br />
Exponent<br />
6 = V notch.<br />
Calculation 2 = Ratiometric.<br />
No. of Alarms 0 = No Alarms<br />
Xducer<br />
1 = dB3<br />
Volume Units 6 = Mil USG<br />
Time Units 4 = Per Day<br />
Measnt. Units 5 = inches<br />
Empty Level 40.0 inches<br />
Minimum Head 0.000 inches<br />
Maximum Head 15.0 inches<br />
Total Enable 1 = On<br />
Totaliser ( R ) 0 = No<br />
Total Multiplier 7 = 1000<br />
Maximum Flow 2.2<br />
Programming is now complete and the unit can now be returned to the run<br />
mode, press CANCEL until Run Mode? Is displayed on the LCD press<br />
ENTER, and the <strong>FlowCERT</strong> LITE will return to Run Mode.<br />