Indicator<br />
P872 Fill Indicator<br />
This parameter determines the rate at which the LCD fill indicator activates.<br />
Default = 10m/min<br />
P873 Empty Indicator<br />
This parameter determines the rate at which the LCD empty indicator<br />
activates. Default = 10m/min<br />
Rate<br />
P874 Rate Update<br />
This parameter determines the way in which the rate is calculated. If set to<br />
continuous (P874=0), then the rate is calculated and displayed<br />
continuously, i.e. any change seen from shot to shot is calculated and<br />
displayed, but if set to use values P874=1(Default) then the values set in<br />
P875 and P876 are used to calculate and display the rate.<br />
P875 Rate Time<br />
This parameter is the period (in seconds) over which the material level rate<br />
of change is averaged before the Rate Value (P877) is updated. If the Rate<br />
Distance (P876) is exceeded before the Rate Time (P875) has expired then<br />
the Rate Value (P877) will be updated immediately. Default = 60sec.<br />
P876 Rate Distance<br />
This parameter is the rate Measurement Units (P104) over which the<br />
material level must change before the Rate Value (P877) is updated. If the<br />
Rate Time (P875) expires before the Rate Distance (P876) is exceeded<br />
then the Rate Value (P877) will be updated immediately. Default = 0.05m<br />
P877 Rate Value<br />
This parameter displays the current rate of change of material level, in<br />
Measurement Units (P104) per minute. It is read only.<br />
P878 Lower Cutoff<br />
This parameter is used to select the minimum Rate to be calculated, and can<br />
be used to eliminate unwanted updates from effects of ripples/waves on the<br />
surface of the material.<br />
Page 111