MED ASSOCIATES INC. <strong>ACTIVITY</strong> <strong>MONITOR</strong><br />
Figure 1-3 - Activity Chamber with Hole Board Insert<br />
• Hole Board Transmitters 1 and 2 connect to the “Transmitter 3” port on the ENV-<br />
520 Controller.<br />
• Hole Board Receivers 1 and 2 connect to the “Receiver 3” port.<br />
• Transmitter 1 gets connected to the “Transmitter 1” port.<br />
• Transmitter 2 gets connected to the “Transmitter 2” port.<br />
• Receiver 1 gets connected to the “Receiver 1” port.<br />
• Receiver 2 gets connected to the “Receiver 2” port.<br />
An easy way to check that the I/R Array Strips are correctly connected is to run a quick<br />
one-minute test experiment. Place an object into the test chamber and confirm that the<br />
red dot on the run-time screen is tracking correctly (i.e., back left in the chamber should<br />
be top left on the screen and front right in the chamber should be bottom right on the<br />
screen).<br />