Standard Operations<br />

5.7.3.2 Wavelength scan reading<br />

64<br />

Cursor mode<br />

1. Press Cursor Mode: Track in the Options menu.<br />

2. The selection for this menu item determines what data are<br />

displayed in the table. Highlight Track or Peak/Valley.<br />

3. Press OK to confirm.<br />

4. Press Return to return to the scan mode.<br />

Integral<br />

The Integral applies to the whole wavelength range of the scan.<br />

1. Press Integral: Off in the Options menu.<br />

2. Highlight On to show the Integral. To find the integral of other<br />

wavelength ranges, change the wavelength range and scan<br />

again.<br />

3. Press OK to confirm.<br />

4. Press Return to return to the scan mode.<br />

Note: The Integral is shown instead of the date on the display.<br />

Note: For the next scan measurement the setting for the Integral will be<br />

On.<br />

Scale & units<br />

1. Press Scale & Units.<br />

2. Highlight the required units (Abs or %T).<br />

3. Highlight Auto or Manual scaling on the graph´s y-axis .<br />

Note: If manual scaling is selected, use the alphanumeric keypad to set the<br />

limits y min. and y max.. The graph is adjusted to display only the values in the<br />

selected range. If automatic scaling is selected, the instrument sets the<br />

limits automatically so that the total range can be displayed.<br />

4. Press OK.<br />

5. Press Return to return to the scan mode.<br />

After the scanning parameters have been selected, the baseline<br />

must be scanned. Changing any of the scanning parameters