Maintenance<br />

7.3 Filter pad maintenance<br />

7.3.1 Filter pad replacement<br />

84<br />

To determine when the filter pad needs to be replaced, inspect the<br />

filter pad every 3–6 months (in a relatively dust-free environment,<br />

this interval can be longer).<br />

1. Remove any cuvettes/sample cells and cell adapter from the<br />

cell compartment.<br />

2. Turn the instrument off.<br />

3. Unplug the power cord.<br />

4. Lift the instrument and check the color of the filter pad. Replace<br />

the filter pad if is dark gray or black.<br />

5. Turn the instrument over and place it on a soft surface.<br />

6. Use a screwdriver (standard or cross-head) to open the<br />

filter grid (Figure 8, item 1).<br />

7. Lift the filter grid (Figure 8, item 1).