Section 8 Troubleshooting<br />

Problem/Display screen Likely Cause Action<br />

Please insert adapter Z.<br />

Color = ***<br />

For measurements with 11mm round<br />

cells the adapter Z is required.<br />

Spectrum locus outside the measuring<br />

range.<br />

Absorbance > 3.5! The measured absorbance exceeds 3.5<br />

Concentration too high!<br />

Error<br />

Selfcheck stopped.<br />

Please check the lamp.<br />

Please close the lid.<br />

Error [xx]<br />

Error<br />

Selfcheck stopped. Please remove<br />

the cuvette<br />

Please close the lid.<br />

Error<br />

Selfcheck stopped.<br />

Hardware error.<br />

Error [x]<br />

Error<br />

Too much ambient light!<br />

Move device into shade<br />

or close the lid!<br />

Calculated concentration is higher than<br />

999999<br />

Self-Check Test stops while starting the<br />

instrument<br />

Self-Check Test stops while starting the<br />

instrument<br />

Electronic defect<br />

The instrument sensors detects too<br />

much ambient light.<br />

Insert the adapter Z into cell<br />

compartment #2.<br />

Press OK.<br />

Dilute the sample or select appropriate<br />

color scale.<br />

Dilute the sample and repeat the<br />

measurement<br />

Dilute the sample and repeat the<br />

measurement<br />

Check the lamp and replace, if<br />

necessary.<br />

Close the lid.<br />

Press Start Again.<br />

Remove the cuvette/sample cell from<br />

the cell compartment.<br />

Press OK.<br />

Contact the manufacturer or a sales<br />

representative and indicate the error<br />

number<br />

Reduce ambient light. (Avoid direct sun<br />

light.)<br />

Close the lid.<br />

Negative result! The calculated result is negative Check the concentration of the sample<br />

No evaluation!<br />

Over measuring range<br />

Under measuring range<br />

Error in the test database / user<br />

database<br />

The measured absorbance is above the<br />

calibration range of the test<br />

The measured absorbance is below the<br />

calibration range of the test<br />

Check the programming<br />

Contact the manufacturer or a sales<br />

representative<br />

Dilute the sample and repeat the<br />

measurement<br />

If possible, select a test with a lower<br />

measurement range or use a<br />

cuvette/sample cell with a longer path<br />

length<br />