Maintenance<br />

WARNING<br />

Burn Hazard. Wait until the lamp cools down. Contact with the<br />

hot lamp can cause burns.<br />

6. Push down on the pressure spring (see instrument label 2).<br />

7. Remove the halogen lamp and the plug panel (see instrument<br />

label 3).<br />

Important Note: Hold the lamp by the fitting only. Avoid touching<br />

the glass, as substances on the skin can bake onto the lamp bulb<br />

and thus accelerate the ageing process of the lamp.<br />

8. Plug a new halogen lamp to the panel.<br />

9. Insert the halogen lamp with the half rounded part pointing up.<br />

10. Press the plug with slight pressure into the direction of the<br />

halogen lamp and push the pressure spring up, so that it will<br />

engage.<br />

11. Fold the fan again, so that it engages.<br />

12. Use a screwdriver to screw the back cover to the instrument.<br />

13. Plug in the power supply.<br />

14. Reset the Lamp History, see section 6.1.5 on page 74.<br />