Advanced Operations<br />

6.1.4 Output checks<br />

6.1.5 Lamp history<br />

74<br />

6. Remove any cuvettes/sample cells from the cell compartment<br />

and press Start.<br />

7. Insert the different filter in the given order one after the other.<br />

Press Next after inserting a filter.<br />

After the last measurement the results are displayed.<br />

8. Press PC & Printer icon to send the data to a USB memory<br />

stick, PC or to a printer.<br />

The files will be stored automatically as CSV file (Comma<br />

Seperated Value). The file name will be formatted as<br />

“Verification.csv“.<br />

If a printer is connected a test printing of the current screen will be<br />

printed.<br />

The Lamp History menu provides the amount of time that the lamp<br />

has been on (Hours).