chromatography. Abdel-Alim <strong>and</strong> Hamielec (41) used a broad MWD PAM<br />

st<strong>and</strong>ard A to create a linear calibration curve that covers the molecular weight<br />

range from 10 3 to 10 7 g/mole. This calibration was used to characterize two<br />

other broad-MWD st<strong>and</strong>ards B <strong>and</strong> C.<br />

The Micropak TSK Gel PW, TSK Gel PWXL <strong>and</strong> Shodex OHpak Q-800,<br />

B-800, <strong>and</strong> KB-800 series are more recently available columns developed for<br />

analyzing the acrylamide polymers <strong>and</strong> other water-soluble polymers in aqueous<br />

SEC. The TSK columns have been evaluated by Barth (25), Alfredson et al. (42),<br />

Sasaki et al. (43), <strong>and</strong> Lin <strong>and</strong> Getman (44). Dhowa Denko (45) reported the SEC<br />

analysis <strong>of</strong> PAM by Shodex OHpak columns. The narrow MWD polyacrylamide<br />

st<strong>and</strong>ards (M w ¼ 1.2 10 4 to 9.0 10 6 g/mole) produced by the American<br />

Polymer St<strong>and</strong>ards Corporation are listed in Table 2. However, some acrylamide<br />

copolymers <strong>and</strong> terpolymers are heterogeneous (40) in terms <strong>of</strong> chemical structure<br />

<strong>and</strong> MW, <strong>and</strong> the st<strong>and</strong>ards having chemical structures similar to the samples are<br />

not commercially available. The absolute MW <strong>and</strong> MWD <strong>of</strong> these polymers are<br />

difficult to determine using conventional SEC with a single refractive index (RI)<br />

detector <strong>and</strong> using narrow MWD st<strong>and</strong>ards for calibration. The on-line dual or<br />

multidetectors were used in an SEC system to solve the above problems.<br />

Kim <strong>and</strong> co-workers (46) developed a methodology for using RI/LALLS<br />

dual detectors to establish the MW calibration curve <strong>and</strong> peak broadening<br />

parameter for a wide range <strong>of</strong> MW for PAM. Lin <strong>and</strong> Getman (44) determined<br />

the absolute MW <strong>and</strong> MWD <strong>of</strong> PAM, HYPAM, acrylamide/acrylic acid<br />

Table 2 Polyacrylamide St<strong>and</strong>ards (Reported by American Polymer St<strong>and</strong>ards<br />

Corporation)<br />

Nonionic, 100% water-soluble powder<br />

Catalog # M w (g/mole) Mp (g/mole) M n (g/mole) IV a (dL/g)<br />

PAAM9000K 9,000,000 6,500,000 4,250,000 14.600<br />

PAAM6000K 5,500,000 3,695,000 2,460,000 10.385<br />

PAAM1000K 1,140,000 725,000 465,300 3.800<br />

PAAM500K 524,000 331,000 209,600 2.250<br />

PAAM350K 367,000 193,400 141,000 1.650<br />

PAAM80K 79,000 50,500 44,400 0.645<br />

PAAM60K 58,400 46,100 36,500 0.545<br />

PAAM20K 21,900 17,300 13,700 0.255<br />

PAAM10K 11,530 7,950 7,600 0.160<br />

a<br />

IV ¼ Intrinsic viscosity in dL/g in 0.05 M sodium sulphate at 308C.<br />

[n] ¼ kM a , a ¼ 0.66, k ¼ 0.000373.<br />