Figure 1 Analysis <strong>of</strong> proteins at very low flow rate. Column, TSKgel G3000SW, 10 mm,<br />

60 cm 7.5 mm; mobile phase, 0.1 M sodium dibasic phosphate, pH 6.8, þ0.1 M sodium<br />

chloride; flow rate, 50 mL/min; detection, 280 nm, UV; temperature, 228C; injection,<br />

75 mL; sample, 5–10 mg each protein.<br />

In terms <strong>of</strong> efficiency, an optimal packing material should exhibit high<br />

performance as well as the appropriate specific resolution, <strong>and</strong> the column<br />

backpressure should be low.The properties <strong>of</strong> silica gel that are important for its<br />

application as aSEC packing material are listed in Table 3. Also listed are the<br />

typical values <strong>and</strong> the range <strong>of</strong> values for each <strong>of</strong> the properties discussed here.<br />

Table 4provides general data for controlled pore glasses, which have been used<br />

extensively for biopolymer analyses but are not available in particle sizes typically<br />

used for HPLC separations. Porous glass is produced from a ternary system <strong>of</strong><br />

© 2004 by Marcel Dekker, Inc.