adjusted such that the pore size distribution obtained is wider than conventional<br />

singleporesizepackings.TheresultantSECcolumncalibrationexhibitsextended<br />

resolving range, comparable to that <strong>of</strong> mixed gel technology, although overall<br />

linearity <strong>of</strong> the calibration curve is somewhat compromised.<br />

3.4 Mechanical <strong>and</strong> Chemical Stability<br />

All packing materials are subject to the development <strong>of</strong> back pressure under flow<br />

conditions. The mechanical stability <strong>of</strong> the gel will determine its maximum<br />

allowable flow rate in operation. The pressure/flow characteristics, as illustrated<br />

inFig.4,revealboththepermeability<strong>of</strong>thepacking,fromtheinitiallinearportion<br />

<strong>of</strong> the graph, <strong>and</strong> the point at which the gel will compress <strong>and</strong> deform. The<br />

Figure 4 Flow rate vs. column pressure measured for a PLgel 5mm, 100 A ˚ ,<br />

300 7.5 mm column, eluant acetone.<br />