28.02.2013 Views

Handbook of Size Exclusion Chromatography and Related ...

Handbook of Size Exclusion Chromatography and Related ...

Handbook of Size Exclusion Chromatography and Related ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Table 4 Typical Eluent Systems for Synthetic Water Soluble Polymers<br />

Type <strong>of</strong> polymer Typical sample Suitable eluent<br />

Nonionic, hydrophilic Polyethylene oxide,<br />

polyethylene glycol<br />

Pure water<br />

Nonionic, hydrophobic Polyvinylpyrrolidone 0.1–0.2 M salt/buffer with 20–50%<br />

organic solvent<br />

Anionic, hydrophilic Sodium polyacrylate,<br />

sodium hyaluronate,<br />

carboxymethyl cellulose<br />

0.1–0.3 M salt/buffer, pH 7–9<br />

Anionic, hydrophobic Sodium polystyrene 0.1–0.3 M salt/buffer, pH 7–9 with<br />

sulfonate<br />

20–50% organic solvent<br />

Cationic, hydrophilic Chitosan, poly-2-vinyl<br />

pyridine<br />

0.3–0.8 M salt/buffer, pH 2–7<br />

Cationic, hydrophobic Polyethyleneimine 0.3–0.8 M salt/buffer, pH 2–7 with<br />

20–50% organic solvent<br />

For nonionic polymers, pure water can <strong>of</strong>ten be used as eluent although a<br />

lowionicstrengthisagoodsafetymeasure<strong>and</strong>addsadegree<strong>of</strong>reproducibilityto<br />

the system. Polyethylene oxide <strong>and</strong> polyethylene glycol are characteristic <strong>of</strong> this<br />

sample category.<br />

For ionic samples it is recommended that salt/buffer systems are used as<br />

eluents. The salts most commonly used are sodium sulfate, sodium nitrate, <strong>and</strong><br />

sodium acetate, because these cause little corrosion to stainless steel column<br />

hardware even at low pH. Ionic strength is varied according to sample type but<br />

generally does not exceed 1.0 Mas increasing salt concentration will promote<br />

hydrophobic interaction. Often abuffer is used to allow pH to be controlled.<br />

Anionic polymers may be eluted using 0.1–0.3 Msalt/buffer at pH 7–9.<br />

Figure 10 shows the analysis <strong>of</strong> polyacrylic acid (sodium salt), which is atypical<br />

example.Polystyrene sulfonate (sodiumsalt) isalsoananionicpolymer,but<strong>of</strong>ten<br />

does not elute under such conditions as it is relatively hydrophobic. Although the<br />

salt/buffer systemis sufficient to suppressthe ionic interaction, adsorption due to<br />

hydrophobic interaction occurs <strong>and</strong> this has to be overcome by introducing some<br />

organic modifier to the mobile phase as shown in Fig. 11. In the case <strong>of</strong> PL<br />

aquagel–OH, methanol is recommended as an organic modifier although with<br />

other packings different solvents may be used (e.g., acetonitrile with TSK PW<br />

columns). The manufacturers’ recommendations on the use <strong>of</strong> organic solvents<br />

withaqueouspackingsshouldalwaysbefollowedcarefullyasthewrongchoice<strong>of</strong><br />

solvent may irreversibly damage the column.<br />

Cationic polymers may be eluted using rather higher salt concentrations,<br />

0.3–1.0 M, <strong>and</strong> pH in the range 2–7. Atypical analysis <strong>of</strong> poly-2-vinyl pyridine<br />

is shown in Fig. 12. As with the anionic samples, if there is ahigh degree <strong>of</strong><br />

© 2004 by Marcel Dekker, Inc.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!