Figure 2 PEO calibration <strong>of</strong> Ultrahydrogel linear column in 50:50 (vol/vol) MeOH–<br />

water with 0.1MLiNO3.<br />

themobilephaseontheelutiontime<strong>of</strong>PEOst<strong>and</strong>ardsfromaUltrahydrogellinear<br />

column is shown in Table 5.<br />

The PEO st<strong>and</strong>ards elute slightly earlier in the 50:50 methanol–water<br />

mixture than in the 20:80 methanol–water mixture. Because the viscosity <strong>of</strong> the<br />

50:50 mixture (1.59 cP at 258C) is higher than that <strong>of</strong> the 20:80 mixture<br />

(1.30 cP), the retention time in the 50:50 mixture theoretically should be longer<br />

than in the 20:80 mixture because <strong>of</strong> higher viscosity or backpressure on the<br />

column. This indicates theUltrahydrogellinearcolumn canswell slightly more in<br />

Figure 3 PEO calibration <strong>of</strong> Shodex KB-80M mixed column in 20:80 (vol/vol)<br />

methanol–water with 0.1 M LiNO3.<br />

© 2004 by Marcel Dekker, Inc.