Figure 11 Absolute MW calibration polyols. Column set: PLgel, 2 MIXED-D þ<br />

500A ˚ þ 100A ˚ ; mobile phase: THF with 250ppm BHT; flow rate: 1.0mL/min; detector:<br />

Waters 410 DRI; column temperature: 408C. Solid line: (B) polystyrene; dashed line: (W)<br />

polyol oligomers, (4) fractions identified by MALDI/MS.<br />

were resolved well enough up to, at least, the pentamer. Two fractions <strong>of</strong> high MW<br />

SEC effluent were collected for MW determination with mass-assisted laser<br />

desorption ionization/mass spectroscopy (MALDI/MS). An absolute MW<br />

calibration curve was then created using these data points. The calibration curve<br />

can be extended to higher MW using a PS calibration by assuming that the ratio <strong>of</strong><br />

MWpolyol :MWPS remains the same at all retention volumes, which indicates<br />

© 2004 by Marcel Dekker, Inc.