Figure 1 Calibration curves<strong>of</strong> SEC columns. Column (A): one typical column; column<br />

set (B): two typical columns; Column (C): one column with a narrow MW range.<br />

Anarrow but adequate MW range is another very effectiveway to increase<br />

theresolutioninSEC,whichisdemonstratedbycolumnCinFig.1.Figure3isan<br />

example<strong>of</strong>separatinglowMWepoxyresinsusingdifferentpore-sizecolumns(6).<br />

ThelowestMWsample,Epikote828,iswellseparatedbyeithertheShodexA801<br />

column (equivalent to a50A ˚ column) or the A802 column (100A ˚ ).However, the<br />

analysis using the A801 column takes less time. The Epikote 1001 sample is<br />

clearly partially excluded by the A801 column. The A802 column gave good<br />

separation at the low MW region; however, the A803 column (10 3 A ˚ )separates<br />

betterinthehigheroligomerarea.TheA803columnisaclearchoicefortheother<br />

© 2004 by Marcel Dekker, Inc.