2 Command System RIGOL<br />

:CALCulate:FFT:HSCale<br />

Syntax<br />

:CALCulate:FFT:HSCale <br />

:CALCulate:FFT:HSCale<br />

Description<br />

Set the horizontal coefficient in FFT operation. This command indirectly sets the FFT<br />

horizontal scale.<br />

Query the current horizontal coefficient in FFT operation.<br />

Parameter<br />

Name Type Range Default<br />

Discrete {1|2|3|4} 1<br />

Note: you can use the :CALCulate:FFT:HSPan command to set the horizontal scale of FFT directly.<br />

Explanation<br />

1: Horizontal Scale=the current FFT sample rate of the screen/20.<br />

2: Horizontal Scale=the current FFT sample rate of the screen/40.<br />

3: Horizontal Scale=the current FFT sample rate of the screen/100.<br />

4: Horizontal Scale=the current FFT sample rate of the screen/200.<br />

Return Format<br />

The query returns 1, 2, 3 or 4.<br />

Example<br />

:CALCulate:FFT:HSCale 2<br />

The query returns 2.<br />