RIGOL<br />

2 Command System<br />

:AUToscale<br />

Syntax<br />

:AUToscale<br />

Description<br />

Enable the auto setting function.<br />

Explanation<br />

This command is not available when the current state of the Pass/Fail function is<br />

“Enable Test”. For details, refer to the :MASK:ENABle command.<br />

The oscilloscope will adjust the vertical scale, horizontal time base and trigger mode<br />

for optimum display of the waveform. Note that to use the auto setting, the<br />

frequency of the signal under test should be no lower than 50 Hz, the duty cycle be<br />

greater than 1% and the amplitude be at least 20 mVpp.<br />

2-2 <strong>DS2000</strong> <strong>Programming</strong> <strong>Guide</strong><br />

2-2