20.01.2015 Views

Programming Guide DS2000 Series Digital Oscilloscope

Programming Guide DS2000 Series Digital Oscilloscope

Programming Guide DS2000 Series Digital Oscilloscope

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

2 Command System RIGOL<br />

:MEASure:VBASe<br />

Syntax<br />

:MEASure:VBASe <br />

Description<br />

Query the base value of the selected waveform.<br />

Parameter<br />

Name Type Range Default<br />

Discrete {CHANnel1|CHANnel2} CHANnel2<br />

Explanation<br />

The unit depends on the unit of the source channel (refer to<br />

the :CHANnel:UNITs command).<br />

Return Format<br />

The query returns the measurement result in scientific notation.<br />

Example<br />

:MEASure:VBASe CHANnel1<br />

The query returns -1.840000e+00.<br />

You can use the following commands to query the statistic maximum, minimum,<br />

current value, average and standard deviation of the base values of the selected<br />

waveform. Sending the following commands will enable the statistic function<br />

automatically.<br />

• :MEASure:VBASe:SMAXimum<br />

• :MEASure:VBASe:SMINimum<br />

• :MEASure:VBASe:SCURrent<br />

• :MEASure:VBASe:SAVerage<br />

• :MEASure:VBASe:SDEViation<br />

<strong>DS2000</strong> <strong>Programming</strong> <strong>Guide</strong> 2-363

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!