2 Command System RIGOL<br />

:TRIGger:PULSe:UWIDth<br />

Syntax<br />

:TRIGger:PULSe:UWIDth <br />

:TRIGger:PULSe:UWIDth<br />

Description<br />

Set the upper limit of the pulse width in pulse trigger and the unit is s.<br />

Query the current upper limit of the pulse width in pulse trigger.<br />

Parameter<br />

Name Type Range Default<br />

Real 2ns to 4s 2μs<br />

Note: when the trigger condition is PGLess or NGLess, the range is from 10ns to 4s.<br />

Explanation<br />

This command is available when the trigger condition (refer to<br />

the :TRIGger:PULSe:WHEN command) is PLESs, NLESs, PGLess or NGLess.<br />

Return Format<br />

The query returns the upper limit of the pulse width in scientific notation.<br />

Example<br />

:TRIGger:PULSe:UWIDth 0.000003<br />

The query returns 3.000000e-06.<br />