2 Command System RIGOL<br />

:WAVeform:STOP<br />

Syntax<br />

:WAVeform:STOP <br />

:WAVeform:STOP<br />

Description<br />

Set the stop position of internal memory waveform reading.<br />

Query the current stop position of internal memory waveform reading.<br />

Parameter<br />

Name Type Range Default<br />

Integer<br />

NORMal: 1 to 1400<br />

MAX: 1 to the number of effective<br />

point currently on the screen<br />

RAW: 1 to the current maximum<br />

memory depth<br />

--<br />

Explanation<br />

For the memory depth, refer to the :ACQuire:MDEPth command.<br />

The setting of the stop position is limited by the current reading mode of the<br />

waveform (refer to the :WAVeform:MODE command).<br />

Return Format<br />

The query returns an integer.<br />

Example<br />

:WAVeform:STOP 200<br />

The query returns 200.<br />