2 Command System RIGOL<br />

:TRIGger:SLOPe:WHEN<br />

Syntax<br />

:TRIGger:SLOPe:WHEN <br />

:TRIGger:SLOPe:WHEN<br />

Description<br />

Select the trigger condition of slope trigger.<br />

Query the current trigger condition of slope trigger.<br />

Parameter<br />

Name Type Range Default<br />

Discrete<br />

{PGReater|PLESs|NGReater|NLESs|PG<br />

Less|NGLess}<br />

PGReater<br />

Explanation<br />

PGReater: you need to specify a time value (refer to the :TRIGger:SLOPe:TLOWer<br />

command). The oscilloscope triggers when the positive slope time of the input signal<br />

is greater than the specified time.<br />

PLESs: you need to specify a time value (refer to the :TRIGger:SLOPe:TUPPer<br />

command). The oscilloscope triggers when the positive slope time of the input signal<br />

is lower than the specified time.<br />

NGReater: you need to specify a time value (refer to the :TRIGger:SLOPe:TLOWer<br />

command). The oscilloscope triggers when the negative slope time of the input<br />

signal is greater than the specified time.<br />

NLESs: you need to specify a time value (refer to the :TRIGger:SLOPe:TUPPer<br />

command). The oscilloscope triggers when the negative slope time of the input<br />

signal is lower than the specified time.<br />

PGLess: you need to specify an upper limit (refer to the :TRIGger:SLOPe:TUPPer<br />

command) and a lower limit (refer to the :TRIGger:SLOPe:TLOWer command) of<br />

time. The oscilloscope triggers when the positive slope time of the input signal is<br />

greater than the specified lower limit and lower than the specified upper limit.<br />

NGLess: you need to specify an upper limit (refer to the :TRIGger:SLOPe:TUPPer<br />

command) and a lower limit (refer to the :TRIGger:SLOPe:TLOWercommand) of<br />

time. The oscilloscope triggers when the negative slope time of the input signal is<br />

greater than the specified lower limit and lower than the specified upper limit.<br />

<strong>DS2000</strong> <strong>Programming</strong> <strong>Guide</strong> 2-467