2 Command System<br />

:ACQuire:AVERages<br />

Syntax<br />

:ACQuire:AVERages <br />

:ACQuire:AVERages<br />

Description<br />

Set the number of averages and the value should be a power function of 2.<br />

Query the current number of averages of the oscilloscope.<br />

Parameter<br />

Name Type Range Default<br />

Integer 2 to 8192 2<br />

Explanation<br />

Use the :ACQuire:TYPE command to select the average acquisition mode. In this<br />

mode, the oscilloscope averages the waveforms from multiple samples to reduce the<br />

random noise of the input signal and improve the vertical resolution. The greater the<br />

number of averages is, the lower the noise will be and the higher the vertical<br />

resolution will be but the slower the response of the displayed waveform to the<br />

waveform changes will be.<br />

Return Format<br />

The query returns an integer between 2 and 8192.<br />

Example<br />

:ACQuire:AVERages 128<br />

The query returns 128.<br />