2 Command System RIGOL<br />

:FUNCtion:WANalyze:ECDiff<br />

Syntax<br />

:FUNCtion:WANalyze:ECDiff<br />

Description<br />

Query the current frame difference.<br />

Explanation<br />

During the analysis based on trace, the oscilloscope compares each frame with the<br />

templet to compute the difference value and normalizes each value using the<br />

maximum one; then, compares the normalized value of each frame with the<br />

threshold selected to judge whether the frame is an error frame. “CurFrame Diff”<br />

(relative difference) is the normalized value of the difference between the current<br />

frame and templet.<br />

During the analysis based on pass/fail mask, the oscilloscope compares each frame<br />

with the mask to compute the difference value and recognizes the frame whose<br />

difference value is equal to or larger than the threshold selected as an error frame<br />

and the corresponding “CurFrame Diff” is 100%; otherwise, the frame is juged as<br />

correct and the “CurFrame Diff” is 0%. Note that there are only two “CurFrame Diff”<br />

values (100% and 0%) under the analysis based on pass/fail mask.<br />

Example<br />

:FUNCtion:WANalyze:ECDiff<br />

The query returns 100.<br />

<strong>DS2000</strong> <strong>Programming</strong> <strong>Guide</strong> 2-217