Linear Trap Turbopump<br />

Forevacuum Pumps<br />

Functional Description<br />

Vacuum System<br />

The <strong>Orbitrap</strong> itself is separated from the UHV chamber by differential<br />

apertures and is evacuated down to 10 -10 mbar by a 210 L/s turbopump<br />

(TMP 3, see Figure 1-17 on page 1-20). All turbopumps are equipped<br />

with TC 100 control units (manufacturer: Pfeiffer).<br />

A separate turbopump provides the high vacuum for the linear ion trap.<br />

It it is mounted to the bottom of the vacuum manifold of the linear ion<br />

trap. For more information, refer to the <strong>LTQ</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong>.<br />

The rotary vane pumps from the linear trap serve as forepumps for the<br />

two smaller turbopumps (TMP 1 and TMP 2). The exhaust hose from<br />

the forepumps is led to the back of the instrument and connects them to<br />

the exhaust system in the laboratory. The forepumps are located on a<br />

small cart in the forepumps cabinet below the linear trap. See<br />

Figure 1-18.<br />

Oil mist filters<br />

Figure 1-18. Forepumps cabinet<br />

Forepumps<br />

To minimize the ingress of pump oil into the exhaust system, the outlets<br />

of the forepumps are fitted to oil mist filters. See page 3-6 on<br />

instructions about returning the collected oil to the forepumps.<br />

The forevacuum pumps (forepumps) of the linear trap are powered by<br />

the power panel of the linear ion trap.<br />

Thermo Fisher Scientific <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong> 1-21