Gas Supply<br />
Functional Description<br />
Gas Supply<br />
Figure 1-19 shows a schematical view of the gas supply in the<br />
instrument. The <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> uses at least two gases for operation:<br />
• Nitrogen,<br />
• Helium, and<br />
Figure 1-19. Schematical view of the gas supply *<br />
* For parts lists of the gas supply, refer to Table 4-10 on page 4-5.<br />
• Argon (optional).<br />
The linear trap requires high-purity (99%) nitrogen for the API sheath<br />
gas and auxiliary/sweep gas. The <strong>Orbitrap</strong> uses nitrogen as collision gas<br />
(bath gas) for the curved linear trap and the HCD collision octapole.<br />
The linear trap requires ultra-high purity (99.999%) helium for the<br />
collision gas.<br />