Functional Description<br />

Printed Circuit Boards<br />

Figure 1-29. Instrument Control board<br />

Diagnostic LEDs<br />

Status LEDs<br />

The diagnostic LEDs listed in Table 1-7 show the status of applied<br />

voltages to the board. The position of the diagnostic LEDs on the board<br />

is indicated by a white rectangle in Figure 1-29.<br />

Table 1-7. Diagnostic LEDs of the Instrument Control board<br />

No. Name Color Description Normal<br />

Operating Condition<br />

LD1 2.5 V Green 2.55 V Input voltage present On<br />

LD2 3.3 V Green 3.3 V Input voltage present On<br />

LD3 5 V Green 5 V Input voltage present On<br />

LD4 -15 V Green -15 V Input voltage present On<br />

LD5 +15 V Green +15 V Input voltage present On<br />

Additionally, the board has four green LEDs that are directly connected<br />

to the micro controller. They indicate the state of the micro controller<br />

and possible error bits and can be used for software debugging. See<br />

Table 1-8 on page 1-39. The position of the status LEDs on the board is<br />

indicated by a white oval in Figure 1-29 on page 1-38.<br />

1-38 <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong> Thermo Fisher Scientific