Table 1-9. Status LEDs of the Power Distribution board<br />

LED green LED orange Information given by orange LED<br />

Vacuum High vacuum failure High vacuum pressure > 10 -8 mbar<br />

Comm. No communication with<br />

instrument control board<br />

CAN bus problem or<br />

instrument control board not<br />

working<br />

System System is not ready FT Electronics switch off or<br />

Vacuum Pumps switch off<br />

Scan Instrument is not scanning<br />

Electr. On Service mode FT Electronics switch off<br />

Vac. Units Ok Vacuum measurement<br />

failure<br />

Pirani <strong>Orbitrap</strong> Ok No function, at present<br />

Vacuum gauge defective<br />

Pirani LT Ok Pirani <strong>LTQ</strong> <strong>XL</strong> failure Control signal < 0.5 V<br />

Ion Gauge ON Penning <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> off Forevacuum > 10 -2 mbar<br />

Ion Gauge OK Penning <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong><br />

failure<br />

Control signal < 0.5 V<br />

LT Vacuum Work <strong>LTQ</strong> <strong>XL</strong> vacuum failure Vacuum forepump <strong>LTQ</strong> <strong>XL</strong><br />

>10 -1 mbar<br />

Vac. 10 -3 mbar<br />

Vac. 10 -5 mbar<br />

Pumps OK Pumps Off Pump down; leakage<br />

Rough P. 1 ON Forepump #1 failure Forepump defective<br />

Turbo P. 1 ON Turbopump #1 failure Turbopump defective/error *<br />

Rotation 1 OK Turbopump #1 failure 80% rotation speed of turbopump<br />

not reached<br />

Turbo P. 2 ON Turbopump #2 failure Turbopump defective/error *<br />

Rotation 2 OK Turbopump #2 failure 80% rotation speed of turbopump<br />

not reached<br />

Heater Off Heater enabled Heater enabled<br />

LAN Conn. Ok LAN connection failure LAN interrupted (Option)<br />

EI ON No function, at present<br />

A System reset System reset has occured<br />

B Micro controller idle<br />

* An error of turbopump 3 is indicated by an LED directly located on the pump controller.<br />

Functional Description<br />

Printed Circuit Boards<br />

Thermo Fisher Scientific <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong> 1-41