Shutting Down the System<br />

Basic System Operations<br />

Shutting Down the System<br />

The <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> system does not need to be shut down<br />

completely if you are not going to use it for a short period of time, such<br />

as overnight or over a weekend. Refer to the topic “Placing the System in<br />

Standby Condition” on page 2-4. This section describes how to shut<br />

down the system for a maintenance or service procedure.<br />

Use the following procedure to shut down the <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong><br />

system:<br />

1. Wait until data acquisition, if any, is complete.<br />

2. Turn off the flow of sample solution from the LC (or other sample<br />

introduction device).<br />

Note For instructions on how to operate the LC from the front panel,<br />

refer to the manual that came with the LC. ▲<br />

3. From the Tune Plus window, choose Control > Off to put the<br />

instrument in Off condition. When you choose Control > Off, all<br />

high voltages are shut off, as are the sheath and auxiliary gas.<br />

4. Put the FT Electronics switch to the OFF position. See Figure 1-7<br />

on page 1-9.<br />

5. Put the Vacuum Pumps switch to the OFF position. See Figure 1-7.<br />

When you place the switch in the OFF position, the following<br />

occurs:<br />

a. All power to the instrument, including the turbomolecular<br />

pumps and the rotary-vane pumps, is turned off.<br />

b. After 30 s, power to the vent valve solenoid of the ion trap is<br />

shut off. The vent valve opens and the vacuum manifold is<br />

vented with nitrogen to atmospheric pressure through a filter.<br />

You can hear a hissing sound as the gas passes through the filter.<br />

6. Leave the main power switch of the <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> in the On<br />

position.<br />

7. During service or maintenance operations that require opening the<br />

vacuum system of the <strong>LTQ</strong> <strong>XL</strong> or <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong>, always put the<br />

main switch (main circuit breaker) to the OFF position. You can<br />

Thermo Fisher Scientific <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong> 2-5