Shutting Down the System<br />
Basic System Operations<br />
Shutting Down the System<br />
The <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> system does not need to be shut down<br />
completely if you are not going to use it for a short period of time, such<br />
as overnight or over a weekend. Refer to the topic “Placing the System in<br />
Standby Condition” on page 2-4. This section describes how to shut<br />
down the system for a maintenance or service procedure.<br />
Use the following procedure to shut down the <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong><br />
system:<br />
1. Wait until data acquisition, if any, is complete.<br />
2. Turn off the flow of sample solution from the LC (or other sample<br />
introduction device).<br />
Note For instructions on how to operate the LC from the front panel,<br />
refer to the manual that came with the LC. ▲<br />
3. From the Tune Plus window, choose Control > Off to put the<br />
instrument in Off condition. When you choose Control > Off, all<br />
high voltages are shut off, as are the sheath and auxiliary gas.<br />
4. Put the FT Electronics switch to the OFF position. See Figure 1-7<br />
on page 1-9.<br />
5. Put the Vacuum Pumps switch to the OFF position. See Figure 1-7.<br />
When you place the switch in the OFF position, the following<br />
occurs:<br />
a. All power to the instrument, including the turbomolecular<br />
pumps and the rotary-vane pumps, is turned off.<br />
b. After 30 s, power to the vent valve solenoid of the ion trap is<br />
shut off. The vent valve opens and the vacuum manifold is<br />
vented with nitrogen to atmospheric pressure through a filter.<br />
You can hear a hissing sound as the gas passes through the filter.<br />
6. Leave the main power switch of the <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> in the On<br />
position.<br />
7. During service or maintenance operations that require opening the<br />
vacuum system of the <strong>LTQ</strong> <strong>XL</strong> or <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong>, always put the<br />
main switch (main circuit breaker) to the OFF position. You can<br />
Thermo Fisher Scientific <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong> 2-5