Functional Description<br />

Printed Circuit Boards<br />

Data Acquisition Analog Board<br />

Table 1-4. Diagnostic LEDs of the Data Acquisition Digital PCI board<br />

Name Color Description Normal Operating Condition<br />

+5 V Green +5 V voltage present On<br />

+3.3 V Green +3.3 V voltage present On<br />

+2.5 V Green +2.5 V voltage present On<br />

Figure 1-28 shows the data acquisition analog board (P/N 206 4150).<br />

Figure 1-28. Data Acquisition Analog board<br />

This board is an add-on board to the mainboard of the internal<br />

computer. See Figure 1-26 on page 1-34. It is used to convert analog to<br />

digital signals for <strong>Orbitrap</strong> experiments, especially for detecting the ions.<br />

The board contains an ADC for the detection of the transient signal,<br />

with a frequency range from 10 kHz to 10 MHz. Three anti-aliasing<br />

filters for the low, middle and high mass range are automatically selected<br />

by the software.<br />

The data acquisition board provides precision timing to control the<br />

acquisition. Events with lower timing requirements on accuracy are<br />

controlled by the linear ion trap.<br />

The diagnostic LEDs listed in Table 1-5 on page 1-37 show the status of<br />

the voltages applied to the board. The position of the LEDs on the<br />

board is indicated by a white rectangle in Figure 1-28.<br />