User Maintenance<br />
General Remarks<br />
General Remarks<br />
Table 3-1. User maintenance procedures<br />
Preventive maintenance must commence with installation, and must<br />
continue during the warranty period to maintain the warranty. Thermo<br />
Fisher Scientific offers maintenance and service contracts. Contact your<br />
local Thermo Fisher Scientific office for more information. Routine and<br />
infrequent maintenance procedures are listed in Table 3-1.<br />
MS Detector Component Procedure Frequency Procedure Location<br />
Analyzer System bakeout If necessary (e.g. after performing<br />
maintenance work on the vacuum system)<br />
page 3-4<br />
Rotary-vane pumps Add oil If oil level is low Manufacturer’s documentation<br />
Change oil Every 3 months or if oil is cloudy or<br />
discolored<br />
Manufacturer’s documentation<br />
Turbomolecular pumps Exchange lubricant reservoir Yearly Manufacturer’s documentation<br />
page 3-6<br />
Recirculating chiller Check cooling fluid level<br />
Check cooling fluid filter<br />
Check air inlet filter<br />
See manufacturer’s documentation Manufacturer’s documentation<br />
page 3-7<br />
To successfully carry out the procedures listed in this chapter, observe<br />
the following rules:<br />
• Proceed methodically<br />
• Always wear clean, lint-free gloves when handling the components<br />
of the API source, ion optics, mass analyzer, and ion detection<br />
system.<br />
• Always place the components on a clean, lint-free surface.<br />
• Always cover the opening in the top of the vacuum manifold with a<br />
large, lint-free tissue whenever you remove the top cover plate of the<br />
vacuum manifold.<br />
• Never overtighten a screw or use excessive force.<br />
• Dirty tools can contaminate your system. Keep the tools clean and<br />
use them exclusively for maintenance and service work at the<br />
<strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong>.<br />
• Never insert a test probe (for example, an oscilloscope probe) into<br />
the sockets of female cable connectors on PCBs.<br />
3-2 <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong> Thermo Fisher Scientific