User Maintenance<br />

General Remarks<br />

General Remarks<br />

Table 3-1. User maintenance procedures<br />

Preventive maintenance must commence with installation, and must<br />

continue during the warranty period to maintain the warranty. Thermo<br />

Fisher Scientific offers maintenance and service contracts. Contact your<br />

local Thermo Fisher Scientific office for more information. Routine and<br />

infrequent maintenance procedures are listed in Table 3-1.<br />

MS Detector Component Procedure Frequency Procedure Location<br />

Analyzer System bakeout If necessary (e.g. after performing<br />

maintenance work on the vacuum system)<br />

page 3-4<br />

Rotary-vane pumps Add oil If oil level is low Manufacturer’s documentation<br />

Change oil Every 3 months or if oil is cloudy or<br />

discolored<br />

Manufacturer’s documentation<br />

Turbomolecular pumps Exchange lubricant reservoir Yearly Manufacturer’s documentation<br />

page 3-6<br />

Recirculating chiller Check cooling fluid level<br />

Check cooling fluid filter<br />

Check air inlet filter<br />

See manufacturer’s documentation Manufacturer’s documentation<br />

page 3-7<br />

To successfully carry out the procedures listed in this chapter, observe<br />

the following rules:<br />

• Proceed methodically<br />

• Always wear clean, lint-free gloves when handling the components<br />

of the API source, ion optics, mass analyzer, and ion detection<br />

system.<br />

• Always place the components on a clean, lint-free surface.<br />

• Always cover the opening in the top of the vacuum manifold with a<br />

large, lint-free tissue whenever you remove the top cover plate of the<br />

vacuum manifold.<br />

• Never overtighten a screw or use excessive force.<br />

• Dirty tools can contaminate your system. Keep the tools clean and<br />

use them exclusively for maintenance and service work at the<br />

<strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong>.<br />

• Never insert a test probe (for example, an oscilloscope probe) into<br />

the sockets of female cable connectors on PCBs.<br />

3-2 <strong>LTQ</strong> <strong>Orbitrap</strong> <strong>XL</strong> <strong>Hardware</strong> <strong>Manual</strong> Thermo Fisher Scientific